This study aims to evaluate the scanning accuracy of scanners equipped with Active Triangulation, Confocal Microscopy, and Multi-Direct-Capture technologies. Department of Prosthetic Dentistry, Faculty of Dentistry, Ege University , 30 patients who applied for a complete removable denture and met the inclusion criteria will have conventional impressions taken. The resulting master model (reference scan) will be scanned using an industrial scanner with high scanning accuracy (E4 lab scanner, 3shape, Denmark) to obtain digital data. Edentulous ridges; Active Triangulation (Medit i700, Seoul, South Korea), Confocal Microscopy (Trios 5, Henry Schein Dental, Gillingham, Kent, UK), and Multi-Direct-Capture technologies (iTero Lumina, Align Technologies, San Jose, California, USA) will be scanned three times using the scanning protocols recommended by the manufacturers and digitized. All digital data obtained will be evaluated using Geomagic (Studio 2015; 3D Systems Inc., Darmstadt, Germany) 3D analysis software to match the reference data and verify its accuracy according to the accepted reference measurement.
See this in plain English?
AI-rewrites the medical criteria so a patient or caregiver can understand them. Always confirm with the trial site.
Accuracy of Intraoral Scanners Compared With Reference Extraoral Scanner
Timeframe: Two months