This retrospective study will evaluate the coronary angiograms of approximately 100 patients who have undergone invasive angiography and fractional flow reserve (FFR). DICOM files of the angiograms will be analyzed on a secure computer within the cardiac catheterization laboratory using the AutocathFFR software to assess if invasive FFR measurements and automated computer analysis of FFR measurements correlate. AutocathFFR measurement per lesion will be compared to the gold standard, invasive FFR value, where an FFR ≤ 0.80 will be considered "positive", while an FFR \> 0.8 will be considered "negative". The AutocathFFR value will be compared to the invasive FFR measurements. The sensitivity and specificity of the AutocathFFR will be calculated, as well as the AutocathFFR accuracy, positive predictive value and negative predictive value per lesion. Device success will be calculated as the ratio of completed versus initiated AutocathFFR index calculations. Usability of the AutocathFFR software will be evaluated using dedicated questionnaires to be completed by the user (cardiologists).
See this in plain English?
AI-rewrites the medical criteria so a patient or caregiver can understand them. Always confirm with the trial site.
To assess the performance (sensitivity, specificity, accuracy) of the AutocathFFR measurements obtained from angiograms for the non-invasive determination of the presence of hemodynamically significant coronary lesions.
Timeframe: 2 months